Unlock Molecular Detail

Precision Spectroscopy and Ion Chemistry

Create stable ion-generation, transport, selection, and trapping workflows for cryogenic, photoelectron, infrared, photofragment, and reaction-kinetics studies.

Home / Applications / Precision Spectroscopy and Ion Chemistry

How Researchers Use Our Products

How Researchers Use Ardara

Explore the instrument challenge, the Ardara solution selected, and the performance reported by each research team.

Used by researchers 2012

University of California, Berkeley and Lawrence Berkeley National Laboratory

Slow Photoelectron Velocity-Map Imaging Spectroscopy of Cold Negative Ions

Solution used
Multiple radio-frequency supplies for hexapole, quadrupole, and octupole ion optics

The cryogenic photoelectron imaging instrument used Ardara RF supplies at several stages: a hexapole guide operated near 3 MHz and 200 Vpp, a quadrupole stage near 1.15 MHz, and an octupole trap with RF amplitudes up to 500 Vpp. These stages guided and trapped cold negative ions before velocity-map imaging. The completed system produced ion temperatures of 10 +/- 2 K and spectral peak widths as narrow as 4 inverse centimeters. This is a strong example of Ardara supplies coordinating multiple ion-optical elements in a precision spectroscopy instrument.

Published by Hock et al. in Review of Scientific Instruments (2012).

Used by researchers 2018

University of Florida

Operation and Performance of a Mass-Selective Cryogenic Linear Ion Trap

Solution used
Quadrupole mass filter upstream of a cryogenic linear ion trap

An Ardara quadrupole mass filter provided initial mass selection before ions entered a cryogenic linear ion trap. The platform produced reproducible spectra from 17 to 295 K and used stored-waveform inverse Fourier transform isolation to select fragile, nitrogen-tagged ions for background-free infrared spectroscopy. This use case connects Ardara mass-selection hardware with low-temperature ion chemistry and action spectroscopy, where clean precursor selection and stable transmission directly affect spectral quality.

Published by Tesler et al. in Journal of the American Society for Mass Spectrometry (2018).

Used by researchers 2021

Korea Advanced Institute of Science and Technology

Observation of the Ponderomotive Effect in Non-Valence Bound States of Polyatomic Molecular Anions

Solution used
RF generators for hexapole and octopole ion guides

Ardara RF generators powered the hexapole and octopole guides used to transport and prepare molecular anions for photodetachment spectroscopy. The authors reported the first experimental demonstration of the ponderomotive effect in non-valence-bound states of polyatomic molecular anions. The case is notable because the Ardara RF hardware supported a fundamental molecular-physics measurement in which stable ion guidance and well-controlled fields were prerequisites for observing a subtle light-matter interaction.

Published by Kang et al. in Nature Communications (2021).

Used by researchers 2022

FELIX Laboratory, University of Amsterdam, and Radboud University

Gas-Phase Spectroscopy of Photostable PAH Ions from the Mid- to Far-Infrared

Solution used
Electron-impact ionizer for forming radical cations

An Ardara electron-impact ionizer generated radical cations of phenanthrene, pyrene, and perylene for infrared multiple-photon dissociation measurements in a 7 T FT-ICR mass spectrometer. Spectra covering approximately 100 to 1700 inverse centimeters were compared with the NASA Ames PAH database and theoretical predictions. The work demonstrates an Ardara ion source contributing to laboratory astrophysics: producing well-defined polycyclic aromatic hydrocarbon ions needed to interpret unidentified astronomical infrared emission features.

Published by Wiersma et al. in Monthly Notices of the Royal Astronomical Society (2022).

Used by researchers 2023

Korea Advanced Institute of Science and Technology

Excited-State Chemistry of the Nitromethane Anion Mediated by the Dipole-Bound States Revealed by Photofragment Action Spectroscopy

Solution used
Radio-frequency quadrupole mass filter for initial mass selection

The experiment used an Ardara RF quadrupole mass filter to isolate nitromethane anions before transfer into a 25 K ion trap for photofragment action spectroscopy. The authors reported the first experimental observation of an excited dipole-bound state of the nitromethane anion and mapped its role in the excited-state reaction pathway. The Ardara filter provided the clean precursor population needed to assign weak, state-specific photochemical features with confidence.

Published by An et al. in Chemical Science (2023).

Co-developed with Ardara 2026

University of Maryland and Ardara Technologies

Development of a Glow-Discharge Ion-Trap Instrument for Measuring Effective Radiative-Association Rate Coefficients

Solution used
Custom glow-discharge ion source, RF-only rectilinear guide with asymptotic rods, deflectors, quadrupole mass filter, detectors, and linear ion trap

The instrument combined a custom Ardara-designed glow-discharge source with ion transport, mass selection, detection, and trapping subsystems for low-temperature ion chemistry. Its first pressure-dependent study examined Ag+ reacting with O2 and placed a lower limit of 1 x 10^-15 cubic centimeters per molecule per second on the effective radiative-association rate coefficient. This coauthored work documents Ardara's role in building a complete experimental chain for a specialized reaction-kinetics problem.

Published by Kisuryna et al.; Randall E. Pedder was an Ardara-affiliated coauthor in The Journal of Physical Chemistry A (2026).

Plan Your Instrument with Ardara

Share your pressure regime, target mass range, RF or DC requirements, geometry, and control constraints. Our team can help match catalog components to your design or define a custom subsystem.