• For RF-only ion guides: Quadrupoles, Hexapoles, Octopoles, and Ion Funnels.
  • Up to 2000 Vpp RF amplitude at nominal frequencies of 100 kHz to 4 MHz with capacitive load.  (Dual RF configurable for two different frequencies.)
  • Oscillates with capacitive loads as high as 1,000 pF, configurable to drive loads as high as 5,000 pF.
  • RF Amplitude can be set manually via a potentiometer on the front panel, or from an external input connection on the rear panel.
  • Self-oscillating RF supply with variable operating frequency.
    • Automatically varies frequency to match capacitive load, eliminating the need for messy matching networks. 
    • Ready to connect up to your vacuum flange with the supplied MHV cables.  Vary the cable length or capacitive load to vary frequency.
  • DC Bias Offset Input supports pulsed modes of operation.
  • Front panel digital voltmeters show peak-to-peak RF voltage for each RF power supply.
  • Vacuum interlock input on back panel to disable high voltage in vacuum fault conditions.
  • Universal AC input (100-240VAC, 50-60 Hz)